VEGA VEGABAR 44 Profibus PA User Manual
Page 26
26
VEGABAR 44 (Profibus PA)
Click in the window "Conditioning“ to
Linearisation
.
In the window "Span“ you first choose in
which unit the adjustment should be carried
out (mbar, psi, kPa). Enter if you want to edit
the percentage or the current value. If you
want to edit the percentage value, enter the
value (e.g. 80 %). Then click to
Save
. The
span adjustment is finished.
Sensor parameter adjustment – Lin-
earisation curves
Click in the window "Instrument data param-
eter adjustment“ to
Conditioning
.
In the window "Zero adjustment“ you choose
in which unit the adjustment should be car-
ried out (mbar, psi, kPa). Then click to
Save
.
The Zero adjustment is finished.
Setup
Click to
Zero
or
Span
to carry out the
respective adjustment.
Note:
The vessel or the cable must be unpressu-
rised for the zero adjustment! For the span
adjustment, it should be possible to generate
a defined pressure (e.g. 80 %) in the vessel
or on the cable.