VEGA VEGASON 54K…56K User Manual
Page 13
VEGASON 54K … 56K
13
Technical data
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770;
temperature 15°C … 35°C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar.
2)
The adjustment time is the time, required by the sensor, to output the correct level after a quick level
change (with max. 10 % deviation).
Accuracy
1)
(typical values under reference conditions, all statements relate to the nominal measuring
range)
Characteristics
linear
Deviation in characteristics including
linearity, reproducibility and
hysteresis (determined by
limit point methods)
< 0.1 %
Linearity
better than 0.05 %
Average temperature coefficient of the
zero signal
0.06 %/10 K
Resolution in general
max. 1 mm
Resolution of the output signal
0.01 % or 1 mm
Characteristics
1)
(typical values under reference conditions, all statements relate to the nominal measuring
range)
Min. measuring range
(between empty and full adjustment)
> 20 mm (recommended > 50 mm)
Ultrasonic frequency (at 20°C)
- VEGASON 54
30 kHz
- VEGASON 55
18 kHz
- VEGASON 56
10 kHz
Measuring intervals
- VEGASON 54
1.0 s
- VEGASON 55
1.5 s
- VEGASON 56
2.0 s
Beam angle at -3 dB emitted power
- VEGASON 54
4°
- VEGASON 55
5°
- VEGASON 56
6°
Influence of the process temperature
1.8 %/10 K, however is compensated by an
integral, dynamic temperature detection
system in the transducer
Influence of the process pressure
negligible within the approved sensor pressure
range
Adjustment time
2)
- VEGASON 54
> 2 s (depending on parameter adjustment)
- VEGASON 55, 56
> 4 s (depending on parameter adjustment)