Wrong orientation to the product, Sensor too close to the vessel wall, Parabolic effects on dished or arched vessel tops – VEGA VEGAPULS 45 Profibus PA User Manual
Page 28
28
VEGAPULS 42, 44 and 45 – Profibus PA
24908-EN-041227
Wrong orientation to the product
Weak measuring signals are generated if the
sensor is not directly pointed at the product
surface. Orient the sensor axis perpendicu-
larly to the product surface to achieve opti-
mum measuring results.
Direct sensor vertically to the product surface
Correct
Incorrect
Ladder
Ladder
Sensor too close to the vessel wall
If the radar sensor is mounted too close to
the vessel wall, strong false echoes can be
caused. Buildup, rivets, screws or weld joints
superimpose their echoes onto the product
i.e. useful echo. Please ensure a sufficient
distance from the sensor to the vessel wall.
If there are good reflection conditions (liquid
medium, no vessel installations), we recom-
mend locating the sensor where there is no
vessel wall within the inner emission cone. For
products in less favourable reflection envi-
ronments, it is a good idea to also keep the
outer emission cone free of interfering instal-
lations. Note chapter "3.1 General installation
instructions“.
Mounting and installation
Parabolic effects on dished or arched
vessel tops
Round or parabolic tank tops act on the radar
signals like a parabolic mirror. If the radar
sensor is placed at the focal point of such a
parabolic tank top, the sensor receives am-
plified false echoes. The optimum mounting
location is generally in the range of half the
vessel radius from the centre.
Less favourable
Mounting on a vessel with parabolic tank top
Advantageous
Unfavourable