6 instrument modification, 1 exchange of adjustment modules, 1 exchange of adjustment mod- ules – VEGA VEGABAR 41 HART User Manual

Page 40

Advertising
background image

40

VEGABAR 41 (HART

®

)

Instrument modification

6 Instrument modification

6.1 Exchange of adjustment mod-

ules

The modular construction of VEGABAR allows
retrofitting, interchange or removal of the
adjustment and display modules. Data previ-
ously saved (e.g. adjustment values) are not
stored in the adjustment module but in an
EEPROM of the electronics module and
therefore readjustment is not necessary. The
connection of the modules is made via a 4-
pole plug connection.

Exchange of the adjustment module

Removal of the adjustment module

•

Separate VEGABAR from the power
supply

•

Loosen the screws on the upper side of
the housing and remove the cover or the
display module

•

Loosen connection cables from the termi-
nals, if necessary pull out the plug con-
nection of the display module

•

Remove the two screws of the adjustment
module

•

Remove adjustment module and pull out
the plug connection

Insertion of an adjustment module

•

Plug the plug connection of the new ad-
justment module into the plug-in socket of
the sensor electronics

•

Fasten the new adjustment module

•

Reconnect the connection cables, if nec-
essary, also the cable from the display
module

•

Close cover or display module of VEGA-
BAR

•

Reconnect VEGABAR to power supply

VEGABAR

Module
without
adjustment

Module for
menu-driven
adjustment

Module for
basic
functions

Display
module