Ambient conditions
A
mbient temperature
-
V
ersion with plug connector
-20 … +85 °C (-4 … +185 °F)
-
V
ersion with cable outlet
-20 … +60 °C (-4 … +140 °F)
S
torage and transport temperature
-
V
ersion with plug connector
-40 … +100 °C (-40 … +212 °F)
-
V
ersion with cable outlet
-40 … +60 °C (-40 … +140 °F)
Process conditions
T
he specifications of the pressure stage and the product temperature are used as an overview.
T
he specifications of the type label are applicable.
P
ressure stage, process fitting
-
T
hread 316L
PN
60
-
T
hread PVDF
PN
10
P
roduct temperature depending on the measuring cell seal
-
FKM (VP2/A)
-20 … +100 °C (-4 … +212 °F)
-
EPDM (A+P 75.5/KW75F)
-40 … +100 °C (-40 … +212 °F)
V
ibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
7)
Electromechanical data
A
ngled plug connector
-
V
ersion
4-
pin according to ISO 4400
-
S
crew terminals for cable cross-section
up to
2.5
mm² (AWG 14)
-
C
able gland
M
16 (
for cable: ø 4.5 … 10 mm)
C
ircular plug connector
4-
pin with screwed connection M12 x 1
C
able outlet
-
L
ength
5
m (16.4 ft)
-
M
in. bending radius
25
mm (with 25 °C/77 °F)
-
D
iameter approx.
6
mm
Voltage supply
O
perating voltage
8 … 30 V DC
P
ermissible residual ripple
U
ss
<
1 V
L
oad
see diagram
7)
T
ested according to the guidelines of German Lloyd, GL directive 2.
24
VEGABAR
14
9 S
upplement
22441
-EN