VEGA VEGABAR 64 Profibus PA User Manual
Page 57
![background image](/manuals/261530/57/background.png)
Process conditions
The specifications to the pressure stage and the product temperature are used as an overview.
The specifications of the type label are applicable.
Pressure stage, process fitting
-
Thread 316L
PN
60
-
Thread Aluminium
PN
25
-
Thread PVDF
PN
10
-
Hygienic fittings 316L
PN
6, PN 10, PN 25, PN 40
-
Flange 316L
PN
16, PN 40,150 lbs, 300 lbs, 600 lbs
-
Flange with extension 316L
without PN specification, PN 16, PN 40 or 150 lbs,
300 lbs, 600 lbs
-
Flange PVDF
PN
16
Product temperature standard version, depending on the meas. cell seal
12)
-
FKM (VP2/A)
-20 … +120 °C (-4 … +248 °F)
-
EPDM (A+P 75.5/KW75F)
-40 … +120 °C (-40 … +248 °F), 1 h: 140 °C/284 °F
cleaning temperature
-
FFKM (Kalrez 6375)
-10 … +120 °C (+14 … +248 °F)
-
FFKM (Chemraz 535)
-30 … +120 °C (-22 … +248 °F)
Product temperature version with extended temperature range, depending on the meas. cell seal
as well as order specification
-
FKM (VP2/A)
-20 … +150 °C (-4 … +302 °F)
-
EPDM (A+P 75.5/KW75F)
-40 … +150 °C (-40 … +302 °F)
-
FFKM (Kalrez 6375)
-10 … +150 °C (+14 … +302 °F)
-
FFKM (Chemraz 535)
-30 … +150 °C (-22 … +302 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
13)
Shock resistance
Acceleration 100 g/6 ms
14)
12)
With process fitting PVDF, max. 100 °C (212 °F).
13)
Tested according to the regulations of German Lloyd, GL directive 2.
14)
Tested according to EN 60068-2-27.
VEGABAR
64 • Profibus PA
57
10 Supplement
28252
-
EN
-081209