5 Setup
Electronics module • VEGASWING 61, 63
30173-EN-130322
5 Setup
5.1 Setup
When replacing the electronics module, make sure that all adjustment
elements are set to the settings of the old electronics module. Hence
the sensor is immediately ready for operation.
In general, all oscillators of series SW60 can be interchanged.
If you want to use an electronics module with a completely different
signal output, you have to carry out the complete setup.
The necessary, suitable operating instructions manual can be down-
loaded from our homepage under Downloads.
The adjustment should be carried out according to the description in
the operating instructions manual.
Note:
Keep in mind that enamelled instrument versions need special
electronics modules. These electronics modules are called SW60E or
SW60E1.