4 Mounting
VEGACAL 63 • For connection to a signal conditioning instrument
30318-EN-130917
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
down to the lowest point of the vessel bottom.
If the instrument is mounted in the filling stream, unwanted false
measurement signals can be generated. For this reason, mount the
instrument at a position in the vessel where no disturbances, e.g. from
filling openings, agitators, etc., can occur.
This applies particularly to instrument versions with a longer probe.
Fig. 6: Inflowing medium
To compensate the normal voltage loss due to sealing materials, you
have to additionally use disc springs for fastening flange screws on
PTFE coated flanges. Tighten the screws moderately with the torque
stated in the technical data.
Inflowing medium
Torque with PTFE plated
flanges