Note:
Makr sure that the housing is not rotated during the electronics
exchange. The connection plug in the housing for the
electronics can then be in another position.
6
Insert the new oscillator carefully. Make sure that the plug
is in the correct position.
7
Insert the two hold screws and tighten them
8
Connect the internal connection cables as follows: red to
terminal 1, black to terminal 2, on VEGASON 64, 65 and 66
connect also the plug connector
9
Screw the housing cover back on
The electronics exchange is now finished.
As a rule, the exchange of the electronics must be
documented internally when used in Ex applications.