VEGA VEGASON series 60 Electronics module User Manual

Page 12

Advertising
background image

Note:
Makr sure that the housing is not rotated during the electronics
exchange. The connection plug in the housing for the
electronics can then be in another position.

6

Insert the new oscillator carefully. Make sure that the plug
is in the correct position.

7

Insert the two hold screws and tighten them

8

Connect the internal connection cables as follows: red to
terminal 1, black to terminal 2, on VEGASON 64, 65 and 66
connect also the plug connector

9

Screw the housing cover back on

The electronics exchange is now finished.

As a rule, the exchange of the electronics must be
documented internally when used in Ex applications.

12

Electronics module - VEGASON series 60