VEGA VEGAFLEX 66 (-200…+400°C) 4 … 20 mA_HART four-wire User Manual
Page 53
1
2
-110°C
(-166°F)
-200°C
(-328°F)
-20°C
(-4°F)
0°C
(32°F)
-40°C
(-40°F)
60°C
(140°F)
80°C
(176°F)
100°C
(212°F)
200°C
(392°F)
100°C
(212°F)
400°C
(752°F)
300°C
(572°F)
Fig. 28: Ambient temperature - process temperature (version -200 … +400 °C/-328 … +752 °F)
1
Ambient temperature
2
Process temperature (depending on the seal material)
Electromechanical data - version IP 66/IP 67
C
able entry
-
D
ouble chamber housing
l
1
x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1
x blind stopper M20 x 1.5; plug M12 x 1 for
VEGADIS
61 (
optional)
or:
l
1
x closing cap ½ NPT, 1 x blind stopper
½ NPT
,
plug M12 x 1 for VEGADIS 61
(
optional)
or:
l
1
x plug (depending on the version), 1 x blind
stopper M20 x 1.5; plug M12 x 1 for VEGADIS
61 (
optional)
S
pring-loaded terminals for wire cross-sec-
tion
2.5
mm² (AWG 14)
Indicating and adjustment module
V
oltage supply and data transmission
through the sensor
I
ndication
LC
display in dot matrix
A
djustment elements
4
keys
P
rotection rating
-
unassembled
IP
20
-
mounted into the sensor without cover
IP
40
M
aterials
-
H
ousing
ABS
-
I
nspection window
P
olyester foil
VEGAFLEX
66 • 4 … 20
mA/HART four-wire
53
10 S
upplement
34174
-
EN