3 wiring plan – VEGA VEGABAR External housing User Manual

Page 11

Advertising
background image

11

5 Connect the sensor to the external housing

External housing • For pressure transmitter VEGABAR

31087-EN-130322

5. Connect the wire ends as described in chapter "Connection plan".

Take note of the numbering.

6. Connect the screen to the internal ground terminal, connect the

outer ground terminal to potential equalisation

7. Tighten the compression nut of the cable entry. The seal ring must

completely encircle the cable

8. Attach the mounting plate again and tighten the screws
The electrical connection of the sensor to the external housing is

finished.
You find the electrical connection of the electronics module in chapter

"Wiring plan" or in the operating instructions manual of the respetive

sensor.

5.3 Wiring plan

Fig. 7: External housing in conjunction with VEGABAR 51, 52, 53, 54, 55

Overview - VEGABAR

series 50