5 Connecting to power supply
VEGAPULS 62 • 4 … 20 mA/HART - four-wire
36504-EN-121011
5 6 7 8
4...20mA
2
1
2
+
( )
(-)
1
Fig. 25: Electronics compartment, double chamber housing
1 Internal connection to the connection compartment
2 For indicating and adjustment module or interface adapter
Information:
The connection of an external indicating and adjustment unit is not
possible with this double chamber housing.
power supply
4...20mA
active
passive
common
IS GND
5
1
/L
/N
PE
2
+
( )
(-)
6 7
8
Fig. 26: Connection compartment with double chamber housing with mains
voltage
Terminal
Function
Polarity
1
Voltage supply
+/L
2
Voltage supply
-/N
5
4 … 20 mA output (active)
+
6
4 … 20 mA output (passive)
+
7
Mass - output
-
8
Function ground with installa-
tion according to CSA
Electronics compartment
Connection compartment