VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - two-wire Approval according to LPR radio standard User Manual

Page 69

Advertising
background image

-

Compression flange DN 80

5 Nm (3.689 lbf ft)

-

Adapter flange DN 100

7 Nm (5.163 lbf ft)

Max. torque for NPT cable glands and Conduit tubes

-

Plastic housing

10 Nm (7.376 lbf ft)

-

Aluminium/Stainless steel housing

50 Nm (36.88 lbf ft)

Input variable

Measured variable

The measured quantity is the distance between
process fitting of the sensor and product surface.
The reference plane is the seal surface on the
hexagon or the lower side of the flange.

3

4

2

1

Fig. 50: Data of the input variable

1

Reference plane

2

Measured variable, max. measuring range

3

Antenna length

4

Useful measuring range

Max. measuring range

35 m (114.8 ft)

Recommended measuring range

up to 20 m (65.62 ft)

Output variable

Output signal

4 … 20 mA/HART

Fulfilled HART specification

7.0

Signal resolution

0.3 µA

Failure signal current output (adjustable)

mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA

Max. output current

22 mA

Starting current

3.6 mA; ≤ 10 mA for 5 ms after switching on

VEGAPULS

61 • Approval according to LPR radio standard

69

11 Supplement

41713

-EN

-120301

Advertising