41

6 Set up with the display and adjustment module

VEGAFLEX 83 • Profibus PA

44223-EN-130910

Warning:

If a linearization curve is selected, the measuring signal is no longer

necessarily linear to the filling height. This must be considered by the

user especially when adjusting the switching point on the limit signal

transmitter.
In the following, you have to enter the values for your vessel, for

example the vessel height and the socket correction.
For non-linear vessel forms, enter the vessel height und the socket

correction.
For the vessel height, you have to enter the total height of the vessel.
For the socket correction you have to enter the height of the socket

above the upper edge of the vessel. If the socket is lower than the up-

per edge of the vessel, this value can also be negative.

+

h

D

- h

Fig. 29: Vessel height und socket correction value
D Vessel height

+h Positive socket correction value

-h Negative socket correction value

Since the adjustment is very comprehensive, the menu points of

Function Blocks 1 (FB1) were put together in a submenu.

Setup - AI FB1