VEGA VEGAFLEX 83 Foundation Fieldbus PFA insulated User Manual
Page 74
74
11 Supplement
VEGAFLEX 83 • Foundation Fieldbus
44227-EN-130910
0,2 m
(7.874
"
)
0,1 m
(3.937
"
)
0,07 m
(2.756
")
1
L
0
-2 mm
(-0.079
")
2 mm
(0.079
")
15 mm
(0.591
"
)
-15 mm
(-0.591
")
Fig. 53: Deviation VEGAFLEX 83 in cable version (coated) in water
1 Dead band - no measurement possible in this area
L Probe length
L
2 mm
(0.079")
0
-2 mm
(-0.079
")
15 mm
(0.591")
-15 mm
(-0.591
"
)
0,35 m
(13.78
")
0,2 m
(7.874
"
)
0,2 m
(7.874
")
0,05 m
(1.969
")
1
1
Fig. 54: Deviation VEGAFLEX 83 in cable version (coated) in oil
1 Dead band - no measurement possible in this area
L Probe length
Repeatability
≤ ±1 mm
Variables influencing measurement accuracy
Temperature drift - Digital output
±3 mm/10 K relating to the max. measuring range or
max. 10 mm (0.394 in)
Additional deviation through electromag-
netic interference acc. to EN 61326
< ±10 mm (< ±0.394 in)