Model 800 Specifications

Performance

Measurement Range

0 to ±100 V DC or peak AC

Measurement Accuracy

Voltage Monitor Output

Better than ±0.1% of full scale

Voltage Display

Better than ±0.1% of reading, ±1 digit
(referred to measured input)

Speed of Response
(10% to 90%)

Less than 3.5 ms for 100 V step

Large Signal Bandwidth
(-3 db)

DC to greater than 100 Hz for 200 V pp

Input Resistance

Greater than 1 X 10

16

W

Input Capacitance

-15

Less than 1 X 10 F

Voltage Monitor

Output

Two BNC outputs provide buffered low-
voltage replica of the measured voltage

Ratio

20:1 of the measured to output voltage
and 1:1 representation

Output Voltage

Less than ±5 mV

Output Noise

Less than 10 mV rms

Output Impedance

Less than 0.5 W

Features

LCD Display Screen with
LED back-light

127 mm x 38 mm (5" x 1.5") screen
displays prompts and voltage
measurements with date and time

Display Resolution

240 x 64 pixels

Hold Mode

Controlled by local pushbuttons or
remotely through HOLD BNC Connector

Zero

Controlled by local pushbuttons or
remotely by applying a momentary (100
ms minimum) TTL low through a
designated BNC Connector

Null Offsets (Auto Null
Adjustment)

Nulling of the input voltage and current is
necessary when the unit is first powered on,
or when the display prompts that an auto
null adjustment is necessary. Placing the
probe in a holster on the side of the unit
and pressing the NULL OFFSETS
pushbutton provides nullification of the
voltage offsets and the “leakage” currents,
which are dependent on environmental
conditions. An audio tone will signal when
the null offsets process is completed.

TEMP/RH Meter Connector

A connector receives an input from an
optional thermo hygrometer (Omega Model
HH311) and enables the Model 800 to
display environmental data on the LCD
screen.

USB Communications Port

4-pin Type B connector

Mechanical

Dimensions

10.2 cm H x 31.6 cm W x 24.1 cm D
(4” H x 12.5” W x 9.5” D)

Weight

2.5 kg (5.5 lb)

BNC Connectors

Two Voltage Monitors (
Hold
Reset

20:1, 1:1)

Ground Receptacle

Threaded ground stud

Power ON/OFF

A momentary push-button

Probe Connector
Location

Front panel

Operating Conditions

Temperature

10°C to 35°C (50°F to 95°F)

Relative Humidity

0 to 80%, noncondensing

Altitude

To 2000 m (6561.68 ft.)

Electrical

AC Line Cord
Receptacle

Selected per geographic destination

Line Voltage

12 V AC, ±10% @ 1 A

DC Connector

2.1 mm DC power plug

Accessories

Operator’s Manual
with software

PN: 23338

Ground Cord

PN: N9044

AC Line Cord

Selected per geographic destination

USB Cable

1.8 m (6 ft) PN: BA103

Probe

Model 800P PN: 17491

Probe Tips

1.27 mm gold plated rounded tip electrode

Additional tips can be customized in various
other sizes to cater to a wide field of
applications; please contact the factory for
more information

Carrying Case

PN: 47032

Note: All specifications measured with a 5-minute warmup time.