Measurement and power solutions – Trek PZD2000A Piezo Driver-Power Amplifier User Manual

Page 2

Advertising
background image

Current Monitor (cont.)

Output Impedance

47 W

Features

High Voltage On-Off

Switch selectable for local or remote. Local:
Individual push-button switches; Remote: TTL
compatible input. TTL High turns off high
voltage. TTL low turns on high voltage.

Dynamic Adjustment

Graduated 1-turn panel potentiometer is used
to optimize the AC response for various load
parameters.

Current Limit/Trip

Switch selectable for either limit or trip. Grad-
uated 1-turn potentiometer is used to adjust
current limit or trip level from 10 to 200 mA

Out of Regulation
Status

Indicator illuminates and BNC provides a TTL
low when required high voltage is not provided
such as during a current limit

Trip Status

Indicator illuminates and BNC provides a TTL
low when high voltage output trips due to cur-
rent trip, detection of fault or removal of cover

Fault Status

BNC provides TTL low when out of regulation
for greater than 500 ms

Mechanical

Dimensions

266 mm H x 482 mm W 655 mm D
(10.5” H x 19” W x 25.8” D)

Weight

24.9 kg (55 lb)

HV Connector

Alden high voltage connector

BNC Connector

Amplifier input, voltage monitor, current
monitor, digital enable, fault/trip status, out of
regulation status

Operating Conditions

Temperature

0°C to 40°C (32°F to 104°F)

Relative Humidity

To 75%, noncondensing

Altitude

To 2000 meters (6561.68 ft.)

Electrical

AC Line Receptacle

Standard three-prong AC line connector

Line Voltage

Factory set for one of two ranges: 104 to 126 V
AC or 180 to 250 V AC, at 48 to 63 Hz

Power Consumption

1000 VA, maximum

Supplied Accessories

Operators’ Manual

PN: 23271

HV Output Cable

PN: 43406

Line Cord

PN: N5011 (104 to 126 V AC)
Contact Factory: (180 to 250 V AC)

Optional Accessories

HV Output Cable

PN: 43406

Measurement and Power Solutions

TM

www.trekinc.com

Model PZD2000A Specifications

Performance

Output Voltage
Range

0 to ±2 kV DC or peak AC

Output Current
Range

0 to ±200 mA DC or ±400 mA peak AC.
Maximum duration for ±400 mA current pulse is
2 ms at 50% duty cycle using a square wave.*

Maximum Power

500 W (real, apparent or reactive). Unit will trip
off if internal power dissipation exceed 500 W

Input Voltage Range

0 to ±10 V DC or peak AC, noninverting

Input Impedance

25 kW, nominal

DC Voltage Gain

200 V/V

DC Voltage Gain
Accuracy

Better than 0.1% of full scale

DC Offset Voltage

Less than ±2 V

Output Noise

Less than 500 mV rms**

Slew Rate (10% to
90%, typical)

Greater than 750 V/µs

Small Signal
Bandwidth (-3dB)

DC to greater than 100 kHz

Large Signal
Bandwidth
(3% distortion)

DC to greater than 60 kHz

Settling Time to 1%

Less than 50 µs for a 2 kV step

Stability

Drift with Time

Less than 50 ppm/hr, noncumulative

Drift with Temp

Less than 100 ppm/°C

Auto Power Limit

Limits internal power dissipation to protect from
overheating

Voltage Monitor

Ratio

01/200th of the high voltage output

DC Accuracy

Better than 0.1% of full scale

DC Offset Voltage

Less than ±2 mV

Output Noise

Less than 5 mV rms**

Output Impedance

47 W

Current Monitor

Ratio

0.025 V/mA

DC Accuracy

Greater than 1% of full scale

Offset Voltage

Less than ±10 mV

Output Noise

Less than 10 mV rms**

Bandwidth (-3 dB)

DC to greater than 5 kHz

*See Automatic Power Limit feature for limitations
**Measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter

Copyright ã 2012 TREK, INC. All specifications are subject to change. 1239/DEC