Model 10/40a specifications – Trek PD04002A Charged Plate Monitor Controller User Manual

Page 2

Advertising
background image

Performance

Features (cont.)

General

Features

Charged Plate Voltage Range

Dimensions

0 to ±55 V DC or peak AC,

122 mm W x 43 mm H x 153 mm D

nominal.

(4.8² W x1.7² H x 6² D).

Weight

Charged Plate Voltage

0.3 kg (0.6 lb).

Measurement Accuracy

0.2% of full scale.

Positive discharge tests,

The Model PD04002A operates

Large Signal Bandwidth (-3 dB)

Negative discharge tests, and

from a +24 V ±10% supply

DC to greater than 200 Hz.

Balance tests can be initialized

@ 0.21 amps.

from a remote location through

Small Signal Bandwidth (-3 dB)

these digital command signals.

DC to 2.5 kHz.

The voltage of the charged plate
is initialized to +55 V or -55 V for

Charged Plate Self-Discharge

discharge time testing or zero

Rate (no incident ion flow)

for balance testing.

Less than 2 V per minute at
55 V for relative humidity up to
85%.

Charged Plate Capacitance

20 pF ± 4 pF. (The capacitance
is independent of charged plate
connecting cable length.)

Unit Operating Conditions

Temperature

5 °C to 35 °C.

Relative Humidity

±

To 85%, noncondensing.

Relative Humidity

To 85%, noncondensing.

TREK, INC. certifies that each Model

Mounting

PD04002A Charged Plate Monitor is

Model PD04002A Monitor

tested and calibrated to specifications

Four (4) M4 mounting screws

using measurement equipment
traceable to the National Institute of

can be used to secure

Standards and Technology or traceable

the monitor assembly from the

to consensus standards. CE compliant.

bottom.

Caution: These instruments are

Charged Plate

designed to make electrostatic

One (1) M4 mounting screw

voltage measurements only. Do not

can be used to secure the

use this instrument to make

charged-plate monitor assembly

measurements of “hard” voltage

from the bottom.

sources or voltage sources which
can deliver currents high enough to

User Connections

cause harmful shocks or personal

Connection block and

injury.

M4 ground lug.

Power

Connection Block

Phoenix Contact Connector.
Connections include:

+24 V DC power input (+),
+24 V DC return (-),
Voltage Monitor (Hi),
Voltage Monitor (Lo),
Digital Reset (plate 0),
Digital Reset (plate +), and
Digital Reset (plate-)

Charged Plate Assembly

High-Temperature Charged

Input Connector

Plate (Trek Model PD04002AP)

BNC connector.

25 mm x 25 mm (1² x 1²).

Cable Length

3 meters.

Capacitance

Voltage Monitor Output V

20 pF 5 pF.

0 to +5 V DC represents a

Operating Range

charged plate voltage of

-60° to 160° C.

-55 V to +55 V (other scale
factors are available).
When the charged plate is at
-55 volts, the voltage at the
Voltage Monitor Output V HI is
0 V. When the charged plate is
at 0 volts, the voltage at the
voltage monitor is +2.5 volts.
When the charged plate is at
+55 volts, the voltage at the
voltage monitor is +5 volts.

Accuracy

0.2% of full scale.

Digital Command Terminals

Plate (O) discharges the floating
plate to zero volts.
Plate (+) precharges the floating

plate to +55 V.

Plate (-) precharges the floating

plate to -55V.

All the above digital command
inputs are at a 10 V DC High
Level. A minimum of 6 mA of
current sink capability is required
to pull the digital command inputs
to ground (0 V).

Certification

Model 10/40A Specifications

Copyright ã 2011 TREK, INC. All specifications are subject to change. 1128/DEC

Model

Charged Plate Monitor

PD04002A

TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK

(fax) • www.trekinc.com • [email protected]

716-438-7555 • 716-201-1804

Advertising