Model 370

Optional Accessories

Probes

Item

Part No.

Item

Part No.

Electrostatic Voltmeter with 1/100 Monitor Output Ratio . . . . . . . . . . . . . . . 370-1

Data Acquisition Module with IEEE-488 Compatible Interface . . . . . . . . . 16073

Electrostatic Voltmeter with 1/200 Monitor Output Ratio . . . . . . . . . . . . . . . 370-2

603RA Full-Rack Mounting Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 603RA

Electrostatic Voltmeter with 1/300 Monitor Output Ratio . . . . . . . . . . . . . . . 370-3

604RA 1/2 Rack Mounting Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 604RA

Electrostatic Voltmeter with 1/600 Monitor Output Ratio . . . . . . . . . . . . . . . 370-6
Electrostatic Voltmeter with 1/1000 Monitor Output Ratio . . . . . . . . . . . . . 370-10

Standard Probe
Model 7000ER (end-viewing). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7000ER

Line Voltage

Miniature Probes

90 to 127 V AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (Part Number)-L

Model 3800E-2 (end-viewing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3800E-2

180 to 250 V AC . . . . . . . . . . . . . . . . . . . . . . . (Part Number - Contact Factory)

Model 3800S-2 (side-viewing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3800S-2

Performance

Features (cont.)

General

Features

Voltage Monitor Output

Dimensions

Measurement Range

A buffered output providing a low-

108 mm H x 223 mm W x 430 mm D

0 to ±3 kV DC or peak AC.

voltage replica of the measured

(4.25" H x 8.75" W x 17" D).

Measurement Accuracy

voltage. 1/200, 1/300, 1/600, 1/1000

Weight

At the Voltage Monitor

5 kg (11 lb).

options available.

(Offset Voltage and

Better than ±0.05% of full scale.

Output Noise specifications vary

Digital Enable

At the Voltage Display

based on output scale factor.)

BNC connector.

Better than ±0.1% of full scale

Scale Factor

Voltage Monitor Output Connector

±1 count, referred to the voltage

1/100th of the measured voltage.

BNC connector.

monitor.

Offset Voltage (at 1/100th scale)

Ground Receptacle

Speed of Response (10% to 90%)

Less than 10 mV.

Binding post.

Less than 50 ms for a 1 kV step.

Output Noise (at 1/100th scale)

Power Requirements

Less than 20 mV rms (measured

Stability

Line Voltage

with the true rms feature of the

Drift with Time

Factory set for one of two ranges:

Hewlett Packard Model 34401A

Less than 150 ppm/hour,

90 to 127 V AC or 180 to 250 V AC,

digital multimeter).

noncumulative.

at 48-63 Hz (specify when ordering).

Output Impedance

Drift with Temperature

Power Consumption

Less than 0.1 W.

Less than 100 ppm/°C.

60 VA, maximum.

Output Current Limit

Probe-to-Surface Separation

Operating Conditions

±10 mA

2 mm ± 1 mm

(recommended).

Temperature

Digital Enable

0 °C to 40 °C.

An open collector, TTL compatible

Relative Humidity

input to enable or disable the

To 85%, noncondensing.

Zero Control

measurement. A TTL high

Altitude

A momentary push-button switch to

will disable the measurement, while a

To 2000 meters

produce zero volts output when the

TTL low will enable the measurement.

Certification and Compliance

probe is coupled to a known zero volt

Data Acquisition Module (optional)

TREK, INC. certifies that each Model

surface.

Provides data output using an IEEE-

370 is tested and calibrated to

Automatic Gain Control

488 compatible interface.

specifications using measurement

The 370 automatically optimizes the

Resolution

equipment traceable to the National

gain of the AC response when

14 bit.

Institute of Standards and Technology.

changing the type of probe

Accuracy

A Certificate of Calibration

being used or when changing the

12 bit.

accompanies each instrument when it

probe-to-surface separation.

Sampling Period

is shipped from the factory.

Programmable from 10 µs to

Voltage Display

Low-Voltage Safety Compliance

30 minutes.

4-character, 7-segment LED display.

(EN 61010-1)

Storage

Range

0 to ±3000 V.

Overvoltage Category

32 Kbytes (16 Kbytes of data words).

Resolution

1 V.

CAT II: Local-level mains,

Acquisition Start Signal

Zero Offset

±1 count, referred to

appliances, portable equipment.

External trigger or IEEE-488 Talk

the voltage monitor.

Pollution Category

Enable or Group Trigger command.

Sampling

Degree 1: Operate in environments

Successive Data Acquisition

Rate

2.5 readings per

where no pollution or only dry,

Internal timer or external trigger.

second.

nonconductive pollution occurs.

Interface Functions

SH1, AH1, T6, L4, SR1, RL1, PP0,
DC1, DT0, C0, E2.

Address
Switch selectable between 0 and 30.

Model 10/40A Specifications

Model 370 Specifications

Copyright Ó 2011 TREK, INC.
All specifications are subject to change.
1122/DEC

All specifications are with a 3800E-2 probe at a probe-to-surface separation of 2 mm unless otherwise noted.

TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK