Model 820 Specifications

Performance

Measurement Range

0 to ±2 kV DC or peak AC

Measurement Accuracy

Voltage Monitor Output

Better than ±0.1% of full scale

Voltage Display

Better than ±0.1% of reading, ±1 digit
(referred to measured input)

Speed of Response
(10% to 90%)

Less than 500 µs for 1 kV step

Large Signal Bandwidth
(-3 db)

DC to greater than 200 Hz for 4 kV pp

Stability

Contacting - Drift with
time at 22 ºC

Less than 6 V/minute, cumulative
(referred to input)

Noncontacting

Better than 100 ppm/˚C

Input Resistance

Greater than 1 X 10

15

W

Input Capacitance

-15

Less than 1 X 10 F

Voltage Monitor

Output

A BNC output provides a buffered low-
voltage replica of the measured voltage

Ratio

1/200th (standard)

Output Current

±5 mA (minimum)

Output Noise

Less than 10 mV rms

Output Impedance

Less than 0.1 W

Front Panel Meter

Voltage Display

3 ½ digit LED display

Range

0 to

V

±1.9 k

Resolution

1 V

Zero Offset

Less than or equal to 1 count

±

Sampling Rate

1 ms between data points

Features

USB Connector

Allows data transfer to a computer with a
sampling rate of 1 ms between data
points (stream data or block data transfer
protocols). PC software can graph the
unit’s output

Reset Button/Connector

A momentary front panel push-button
switch or rear panel external TTL input
signal initiates a reset function.

Digital Enable

A TTL compatible input to enable or disable
the unit’s high-voltage measurement. A TTL
high will disable while a TTL low will enable
the measurement.

.

Mechanical

Dimensions

10.2 cm H x 22.9 cm W x 33 cm D
(4” H x 9” W x 13” D)

Weight

1.8 kg (4 lb)

BNC Connectors

Voltage Monitor
Digital Enable
Reset

USB Port

Allows data transfer to a computer with a
sampling rate of 1 ms between data points

Ground Receptacle

Threaded ground stud

Power ON/OFF

A momentary push-button

Probe Connector
Location

Front panel

Operating Conditions

Temperature

10°C to 35°C (50°F to 95°F)

Relative Humidity

0 to 75%, noncondensing

Altitude

To 2000 m (6561.68 ft.)

Electrical

AC Line Cord
Receptacle

A universal line PFC-type wall cube provides
input power to the ESVM

Line Voltage

24 V DC, ±5% @ 1 A

DC Connector

2.1 mm DC power plug

Supplied Accessories

Operator’s Manual
with software

PN: 24003

Ground Cord

PN: N9082

AC/DC Universal
Power Cube

PN: L5190 (
264 V AC)

Universal power cube at 90 V to

Probe

Model 820P

Probe Tip

Ceramic tip of 0.8 mm diameter

Optional Accessories

Probe

Model 820P

Probe Tips

Ceramic tip of 0.8 mm

Additional tips can be customized in various
other sizes as small as 100 µm to cater to a
wide field of applications; please contact the
factory for more information

Note: All specifications measured with a 5-minute warmup time.

Copyright ã 2012 TREK, INC. All specifications are subject to change. 1234/DEC