Model 10/40a specifications, Model 668b specifications, Output general features (cont.) – Trek 668B High-Voltage Reference Supply User Manual

Page 2: Mechanical performance features certification, Model 668b accessories, Model 668b ordering information

Advertising
background image

Output

General

Features (cont.)

Mechanical

Performance

Features

Certification

Output Voltage Range

Line Supply

Memory

Adjustable from 0 to +3 kV DC or

90 to 126 V AC or 180 to 250 V AC ±10%,

Nine (9) programmable memory locations

0 to -3 kV DC.

at 50-60 Hz (specify when ordering).

for frequently used output voltage and
current settings. A voltage setting

Output Voltage Resolution

Operating Conditions

(including polarity) and a current value

Temperature

100 mV.

can be stored in each memory location.

5 °C to 35 °C.

Relative Humidity

Output Current Range

To 75%, noncondensing.

0 to ±5 mA DC.

Dimensions

104 mm H x 438 mm W x 394 mm D,
(5.09" H x 17.25" W x 15.51" D).

Output Voltage Accuracy

Front Panel Display

Better than 0.015% of full scale.

A sixteen (16) character LED display that

Weight

indicates the output voltage, the current

6.9 kg (15.2 lb).

Zero Offset Voltage

limit/trip setting, and the memory

Less than 50 mV.

location.

High-Voltage Output Connector

MHV connector.

Output Noise

Front Panel Voltage Control

Less than 15 mV rms (measured using

High Voltage On/Off

Digital Enable Connector

the true rms feature of the Hewlett

Individual push buttons.

BNC connector.

Packard Model 34401A digital

Voltage Selection

multimeter).

Up and down push buttons that

AC Line Receptacle

independently control each of the

Standard three-prong AC line connector

Regulation

5 digits of the displayed voltage.

with integral fuse holder.

Line

Polarity

Better than 0.0025% for a line change

Push-button selectable.

from 90 V to 126 V AC or 180 V to

Zero

250 V AC.

Push button that commands the output

Load

to zero volts.

TREK, INC. certifies that each Model 668B is

Better than 0.0025% for a load change

tested and calibrated to specifications using

from 0 to 5 mA.

Current Limit/Trip

measurement equipment traceable to the

Current LIMIT or current TRIP is selected

National Institute of Standards and

Stability

using a single push-button switch. The

Technology or traceable to consensus

Drift with Time

value is adjustable from 1 mA to 5 mA

standards. A Certificate of Calibration

Less than 2 ppm/month,

with push buttons. A LIMIT or TRIP

accompanies each instrument when it is

noncumulative.

indicator will flash when the instrument is

shipped from the factory.

Drift with Temperature

in a current limit or trip condition.

Less than 25 ppm/°C.

Current Limit/Trip Resolution

1 mA.

Current Limit/Trip Accuracy

Better than 10% of setting.

Digital Enable Feature

An open collector, TTL compatible input
to turn on and off the high-voltage output.
A TTL high will turn off the high voltage.
A TTL low will turn on the high voltage.

Model 10/40A Specifications

All specifications are with no load unless otherwise noted.

Model 668B Specifications

Copyright ã 2011 TREK, INC.
All specifications are subject to change.
1152/DEC

Model 668B

Accessories

Item

Part No.

Item

Part No.

HV Reference Supply (90 to 126 V AC) . . . . . . . . . . . . 668B-L

Operator's Manual . . . . . . . . . . . . . . . . . . . . . . . 23070

HV Reference Supply (180 to 250 V AC) . . . . . . . . . . . . 668B-H

MHV Mating Plug . . . . . . . . . . . . . . . . . . . . . . . . B3005
Line Cord (90 to 126 V AC) . . . . . . . . . . . . . . . . . . N5002
Line Cord (180 to 250 V AC). . . . . . . . . . . . . Contact factory
Model 604RA Rack Adapter (optional) . . . . . . . . . . . . C4102

Model 668B Ordering Information

TREK, INC. 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK

585-798-3140 • 585-798-3106 (fax) • www.trekinc.com • [email protected]

Advertising