Model 541A Specifications

Performance

Measurement Ranges

Model 541A-1

0 to ±1 kV DC or peak AC

Model 541A-2

0 to ±100 V DC or peak AC

Speed of Response (10%
to 90%)

Less than 50 ms for a ±1 kV step (541A-1)

Less than 50 ms for a ±100 V step (541A-2)

Accuracy

1% of full scale

Resolution

0.1%

Monitor Output

Model 541A-1

1/100th of the measured voltage

Model 541A-2

1/10th of the measured voltage

Output Noise

Less than 30 mV rms*

Output Impedance

47 W

Features

Alarms

Activated if measured voltage exceeds
preset threshold limits; positive/negative
limits may be programmed separately

Visual

Front-panel LEDs illuminate when
thresholds are reached

Audible

ON/OFF programmable pulsating or
continuous tone. (+) and (-) alarms have
different tone rates for the pulsating tone
selection

Alarm Relay Output

Form C relay contact with barrier strip
terminals; changes state when voltage
thresholds are met or exceeded

Alarm Digital Output

TTL output with a TTL low (0 to 0.8 V) as
the alarm “ON” status; TTL high (2.5 to 5.0
V) indicates a normal condition

Reset Button

Resets alarms and peak hold to zero

Zero Control

Adjustable to produce zero volts when
probe coupled to a known zero voltage
source

Alphanumeric LCD
Display

20 character by four line (20x4) LCD
displays the present voltage and holds the
most positive and most negative values

Ground Receptacle

Banana jack

Serial Port and USB Port

Provides control of specific functions and
acquires sensor data utilizing Trek software
and a PC connected to the RS-232 serial
port or the USB Type B port (connectors are
on back panel)

Current Output

Provides a current of 4 mA to 20 mA
representing -1 kV to +1 kV (541A-1) and
-100 V to +100 V (541A-2)

Features (cont.)

Menu ↑Buttons

Select and program menu options - the

set the alarm threshold voltages, alarm
conditions and alarm reset type

Mechanical

Dimensions

97 mm H x 152 mm W x 204 mm D
(3.8” H x 6” W x 8” D)

Weight

0.77 kg (1.7 lb.)

Operating Conditions

Temperature

15°C to 35°C (59°F to 95°F)

Relative Humidity

5% to 85% RH, non-condensing

Altitude

To 2000 m (6561.68 ft.)

Electrical

Power

15 V DC ±20%, 800 mA, power bus or
AC/DC adapter with a 2.1 mm DC plug.

Power ON/OFF

Rear panel switch

Supplied Accessories

PN: 24005 Operator Manual (with software CD), PN: N9056 6P/4C
Plug, PN: BA108 Serial Cable, PN: N9044 Ground Cord

Optional Accessories

AC/DC Adapter

PN: F5054R

Walking Test Adapter - Round Body Probe CN: 1K037

Walking Test Adapter - Square Body Probe

CN: 1K038

Probes

Probe

Miniature electrostatic field chopper probe

End-view orientation
(round body)

Model No: 541PR-E. Aperture 1.6 mm (0.06”)
diameter. Dimensions 9.7 mm dia x 50 mm L
(0.38” dia. x 2” L)

Side-view orientation
(round body)

Model No: 541PR-S. Aperture 2.3 mm (0.09”)
diameter. Dimensions 9.7 mm dia x 53 mm L
(0.38” dia x 2.1” L)

Side-view orientation
(square body)

Model 541P-S. Aperture 2.5 mm (0.1”)
diameter. Dimensions 5.6 mm sq x 50 mm L
(0.22” sq x 2” L)

Probe Cable Length

3 meters

nominal

(9.8 ft.)

Probe-to-Surface
Separation Distance

2.5 mm ±1 mm (recommended)

Copyright ã 2013 TREK, INC. All specifications are subject to change. 1340/JRB

End View Probe Aperture
1.6 mm

50 mm

Side View Probe Aperture
2.3 mm

50 mm

50 mm

Side View Probe Aperture
2.5 mm

5.6 mm

Measurement and Power Solutions

TM

*Measured using the true rms feature of the HP Model 34401A digital multimeter