PERFORMANCE

FEATURES

GENERAL

PROBE

NOTE: The Model 706B is
designed to make electrostatic
voltage measurements only!

Certification

Measurement Range

Range Switch

Power Requirements

0 to +1 kV or -1 kV DC

A two-position switch is used to

Two (2) 9-volt alkaline batteries.

(switch selectable).

select the measurement range of
either

0 to +1 kV or 0 to -1 kV.

Operating Time

Accuracy

10 hours after battery replacement.

Better than 0.5% of full scale.

TEST/READ Switch

In the TEST position, if the digital

Operating Conditions

display indicates a value lower than

Temperature

1100, the batteries should be

5 °C to 35 °C.

Relative Humidity

replaced. In the READ position, the

Recommended Probe-to-Surface

To 85%, noncondensing.

measurements are displayed.

Separation Distance

1 mm to 5 mm.

Voltage Display

Dimensions

3½ digit liquid crystal display.

39 mm H x 130 mm W x 134 mm D,

Trek Model 701P-76S Probe

Range

(1.6² H x 5.2² W x 5.3²L).

Dimensions

0 to ±1000 V.

8.7 mm diameter x 70 mm L

Resolution

Weight

(0.35² diameter x 2.75² L).

1 V.

455 g (1 lb), with batteries.

Zero Offset

Caution: The metallic components of

Less than ±1 count.

Ground Receptacle

the probe are driven to the measured

Sampling Rate

The g

reen banana jack must be

surface voltage level with a current

3 readings per second.

connected to ground to maintain

capability of ±0.5 milliamperes. These

measurement accuracy

.

metallic surfaces should not be
connected to ground or touched during

Power On/Off

operation.

A two-position toggle switch.

Cable Length

1830 mm (72").

To ensure safety, the Model 706B
should never be used to perform

TREK, INC. certifies that each Model 706B

Aperture Size (oval shaped)

measurements of “hard” voltage

is tested and calibrated to specifications
using measurement equipment traceable

3.18 mm W x 6.35 mm L

sources or voltage sources

to the National Institute of Standards and

which can deliver currents high

(0.125² W x 0.25² L).

Technology.

enough to cause harmful shocks
or personal injury.

Aperture Orientation

RoHS and CE compliant.

Side viewing.

Model 706B Included Accessories

Certification (cont.)

Item

Part No.

Portable Electrostatic Voltmeter . . . . . . . . . . . . . . . . . . . . 706B

Operator's Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23389

Two (2) (9 V) Alkaline Batteries . . . . . . . . . . . . . . . . . . . F1003R

Ground Patch Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . N9044

RoHS compliant

Low Voltage Safety Compliance (IEC 61010)

Overvoltage Category

CAT I: Peripheral level outputs (less than 60 volts).

Pollution Category

Degree 1: Operate in environments where no
pollution or only dry, nonconductive pollution occurs.

Model 10/40A Specifications

Model 706B Specifications

Copyright Ó 2011 TREK, INC.
All specifications are subject to change.
1122/DEC

All specifications are at a probe-to-surface separation of 3 mm unless otherwise noted.

TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK

(fax) • www.trekinc.com • [email protected]

716-438-7555 • 716-201-1804

Advertising