Measurement and power solutions, Models p0621p and n specifications – Trek P0621 High-Voltage Power Amplifier User Manual

Page 2

Advertising
background image

Features

High Voltage On/Off

Local

Individual push-button switches

Remote

TTL high turns OFF the high voltage; TTL low
turns on the high voltage

Dynamics
Adjustments

Graduated 1-turn potentiometer used to
optimize the AC response for various load
parameters

Current Limit/Trip

Switch selectable for limit or trip. Graduated 1-
turn potentiometer adjusts from 0 to 20 mA

Out of Regulation

LED illuminates and BNC provides a TTL low
when P0621 fails to produce HV output such as
during a current limit

Trip Status

LED illuminates and BNC provides a TTL low
when HV is disabled due to the output current
exceeding the current trip level, a high voltage
fault is detected or the top cover is removed

Fault Status

BNC provides a TTL low when P0621 is out of
regulation for greater than 500 ms

Mechanical

Dimensions

222 mm H x 432 mm W x 584 mm D
(8.75” H x 17” W x 23” D)

Weight

24.9 kg (55 lb)

HV Connector

Caton High Voltage Connector

BNC Connectors

Amplifier input, voltage monitor, current
monitor, remote HV ON/OFF, out of regulation,
fault/trip status

Operating Conditions

Temperature

0°C to 40°C (32°F to 104°F)

Relative Humidity

To 75%, noncondensing

Altitude

To 2000 meters (6561.68 ft.)

Electrical

Line Voltage

Factory Set for one of two ranges:
90 to 127 V AC or 180 to 250 V AC,
either at 48 to 63 Hz

AC Line Receptacle

Standard 3-prong with integral fuse holder

Power Consumption

1000 VA, maximum

Supplied Accessories

Operators’ Manual

PN: 23147

HV Output Cable

PN: 47067

Line Cord (for 104 to
127 V AC)

PN: N5011

Line Cord (for 180 to
250 V AC)

Selected per geographic destination

Copyright ã 2012 TREK, INC. All specifications are subject to change. 1251/JRB

Models P0621P and N Specifications

Performance

Output Voltage
Positive Polarity
(P0621P)

0 to +30 kV DC or peak AC

Output Voltage

Negative Polarity
(P0621N)

0 to -30 kV DC or peak AC

Output Current

0 to ±20 mA

Input Voltage Range
(P0621P)

0 to +10 V DC or peak AC

Input Voltage Range
(P0621N)

0 to -10 V DC or peak AC

Input Impedance

50

kW, nominal

DC Voltage Gain

3000 V/V

DC Voltage Gain
Accuracy

Better than 0.1% of full scale

Offset Voltage

Less than ±4 V

Output Noise

Less than 5 V rms*

Slew Rate
(10% to 90%, typical)

Greater than 350 V/µs

Large Signal
Bandwidth
(1% distortion)

DC to greater than 3.5 kHz

Small Signal
Bandwidth (-3dB)

DC to greater than 25 kHz

Automatic Power
Limit

Automatically limits the internal power
dissipation to protect the Models P0621P and
P0621N from overheating

Voltage Monitor

Ratio

1/3000th of the HV output signal

DC Accuracy

Better than 0.1% of full scale

Offset Voltage

Less than ±2 mV

Output Noise

Less than 20 mV*

Output Impedance

47

W

Current Monitor

Ratio

0.5 V/mA

DC Accuracy

Better than 1% if full scale

Offset Voltage

Less than ±10 mV

Output Noise

Less than 30 mV rms*

Bandwidth

DC to greater than 10 kHz

Output Impedance

47

W

Measurement and Power Solutions

TM

www.trekinc.com

*Measured using the true rms feature of the HP Model 34401A digital multimeter

Advertising