Measurement and power solutions, Model 603 specifications – Trek 603 Piezo Driver-Power Amplifier User Manual

Page 2

Advertising
background image

Features (cont.)

Digital Enable

An input providing a connection for a TTL
compatible signal to turn on/off the HV output

Load Range Switch

Slide switch to select high or low capacitive
loads (more than 150 pF or less than 150 pF)

Dynamic Adjustment

Graduated 1-turn panel potentiometer is used
to optimize the AC response for various load
parameters.

Mechanical

Dimensions
Single Channel

Instrument

Double Channel

Instrument

222.3 mm H x 108 mm W 381 mm D
(8.75” H x 4.25” W x 15” D)

433.8 mm H x 108 mm W 381 mm D
(17” H x 4.25” W x 15” D)

Weight
Single Channel

Instrument

Double Channel

Instrument

4.3 kg (9.4 lb)

8.6 (18.8 lb)

HV Connector

SHV High Voltage Connector

BNC Connectors

Power Switch, Amplifier Input, Voltage Monitor,
Current Monitor, High Voltage ON/OFF, Digital
Enable

Operating Conditions

Temperature

0°C to 40°C (32°F to 104°F)

Relative Humidity

To 85%, noncondensing

Altitude

To 2000 meters (6561.68 ft.)

Electrical

Line Voltage

Factory Set for one of two ranges:
90 to 127 V AC or 180 to 250 V AC,
either at 48 to 63 Hz

AC Line Receptacle

Standard 3-prong with integral fuse holder

Power Consumption

125 VA, maximum

HV Cable

2 m, 66 pF per foot

Supplied Accessories

Operators’ Manual

PN: 23166

HV Output Cable

PN: 43874

Line Cord
Spare Fuses

N5002
Selected per geographic destination

Optional Accessories

19” Rack Mount Kit
Half-Rack Mount Kit
Dual Instrument Rack
Mount Kit

Model 603RA
Model 604RA (3 ½” Buckeye)
Model 603RA-2

Copyright ã 2012 TREK, INC. All specifications are subject to change. 1230/DEC

Model 603 Specifications

Performance

Available Output
Voltage Ranges

0 to ±125 V DC or peak AC
0 to -250 V DC or peak AC
0 to +250 V DC or peak AC

Output Current

±40 mA DC
±80 mA peak AC, for less than 1 ms

DC Voltage Gain

50 V/V (a gain of 25 V/V is available)

DC Voltage Gain
Accuracy

Better than 0.1% of full scale

DC Offset Voltage

Less than ±500 mV

Output Noise*

Less than 20 mV rms

Slew Rate
(10% to 90%, typical)

Greater than 100 V/µs

Large Signal
Bandwidth
(5% distortion)

DC to greater than 150 kHz

Settling Time to 1%

Less than 5 µs for a 100 V step

Stability

Drift with Time

Less than 100 ppm/hr, noncumulative

Drift with

Temperature

Less than 25 ppm/°C

Amplifier Input

Input Voltage Range

0 to ±10 V DC or peak AC, noninverting

Input Impedance

10 W, nominal

Voltage Monitor

Ratio

1/25th of the high voltage output

DC Accuracy

Better than 0.1% of full scale

AC Accuracy

Calibrated using a Ross Model VD30-4.1-BD-
KC-ALU high voltage divider

DC Offset Voltage

Less than 5 mV

Output Noise

Less than 5 mV rms*

Output Impedance

0.1 W

Current Monitor

Ratio

0.1 V/mA

DC Accuracy

Greater than 1% of full scale

Offset Voltage

Less than 10 mV

Output Noise

Less than 10 mV rms*

Output Impedance

0.1 W

Features

Output Voltage
Configuration

Factory set for 0 to ±125 VDC or peak AC.
Other ranges available.

*Measured using the true rms feature of the Hewlett Packard 34401A digital multimeter

Measurement and Power Solutions

TM

Advertising