Appendix A - Troubleshooting

Problem

Cause / Possible Solution

ESR value when probes

are shorted is not close

to 0Ω

Perform a probe compensation.

Seems to be stuck

“Measuring

capacitance

It is possible that one (or more) of the 4-wire

probe connections has become broken. Check all

connections.

Display shows

Removing trace

charge

This message is displayed if the Atlas ESR has

detected that the attached capacitor may be

exhibiting “Soakage” or “Dielectric

Absorption”, this is quite normal. The

instrument then ensures that the capacitor is very

well discharged and helps to prevent voltage

developing across the capacitor after the normal

discharge procedure has completed.

Display shows

Auto discharge

taking too long!

The unit attempts to remove charge from the

capacitor using a controlled discharge procedure.

If this takes longer than 60 seconds then the

discharge process will be aborted. It is

recommend that you safely discharge the

capacitor manually and try analysis again.

Display shows

Warning! V=132V

Safely discharge

If the voltage across the capacitor is greater than

50V then the Atlas ESR will not attempt to

discharge the capacitor, please safely discharge

the capacitor manually.

Display shows

Self Test Fail

Code 2

It is possible that a hardware failure has

occurred, please contact Peak Electronic Design

Limited for assistance.