3B Scientific Electron Deflection Tube D User Manual

3b scientific® physics

Advertising
background image


3B SCIENTIFIC® PHYSICS



1

Electron-Beam Deflection Tube D 1000651


Instruction sheet

11/12 ALF

5

7

2

1

-

4

6

2
1

1
2

10 9 8 7 6 5 4 3 2

3

1 Fluorescent screen
2 Lower deflection plate
3 Boss with 4-mm plug for

connecting deflection plates

4 Electron gun
5 4-mm sockets for connecting

heater supply and cathode

6 4-mm plug for connecting

anode

7 Upper deflection plate

1. Safety instructions

Hot cathode tubes are thin-walled, highly

evacuated glass tubes. Treat them carefully as

there is a risk of implosion.

•

Do not subject the tube to mechanical stresses.

•

Do not subject the connection leads to any

tension.

•

The tube may only be used with tube holder

D (1008507).

If voltage or current is too high or the cathode is

at the wrong temperature, it can lead to the tube

becoming destroyed.

•

Do not exceed the stated operating parameters.

•

Only change circuit with power supply

equipment switched off.

•

Only exchange tubes with power supply

equipment switched off.

When the tube is in operation, the stock of the

tube may get hot.

•

If necessary, allow the tube to cool before

dismantling.

The compliance with the EC directive on elec-

tromagnetic compatibility is only guaranteed

when using the recommended power supplies.

2. Description

The electron-beam deflection tube is intended

for investigating the deflection of electron beams

in electrical and magnetic fields. It can be used

to estimate the specific charge of an electron
e

/m and to determine the electron velocity v.

The electron-beam deflection tube comprises an

electron gun which emits a narrow, focussed

ribbon of cathode rays within an evacuated,

clear glass bulb. A tungsten 'hairpin' filament hot

cathode is heated directly and the anode takes

the form of a cylinder. The deflection of rays can

be achieved electrostatically by means of a built-

in plate capacitor formed by the pair of deflec-

tion plates or magnetically with the help of the

Helmholtz coils D (1000644) magnetically. The

cathode rays are intercepted by a flat mica

sheet, one side of which is coated with a fluo-

rescent screen and the other side of which is

printed with a centimetre graticule so that the

path of the electrons can be easily traced. The

mica sheet is held at 15° to the axis of the tube

by the two deflecting plates.


Advertising