3B Scientific DC Power Supply, 450 V (115 V, 50__60 Hz) User Manual
Page 6
2
2. Description
The DC power supply 450 V is designed to supply
power for electrostatics experiments, in particular
experiments using an electrometer (U8531408).
It possesses two DC outputs and an AC output, all of
which are protected against short circuits.
The power supply
with the item number U8521400-
115 is for operation with a mains voltage of 115 V
(±10%), and the transformer with the item number
U8521400-230 unit is for operation with a mains
voltage of 230 V (±10%).
3. Technical data
Mains voltage:
see rear of housing 230/115
V AC, 50/60 Hz, 10 VA
Fuses:
Internal quick-blow fuses
5x20, 250 V AC,
T800 mA for 115 V AC
T400 mA for 230 V AC
Output 1:
Voltage:
0 – 450 V DC , continuously
adjustable
Max. current:
10 µA
Internal resistance: 1 M
Ω
Output 2:
Voltage:
1.2 – 12 V DC, continuously
adjustable
Max. current:
100 mA
Output 3:
Voltage:
0 – 12 V AC, continuously
adjustable
Max. current:
10 mA
Terminals:
4 mm safety sockets
Dimensions: 250x100x160
mm³
approx.
Weight:
0.8 kg approx.
4. Operation
4.1 General information
•
The AC and DC outputs cannot be used at the
same time
•
Before connecting the plug to the power supply,
set the voltage selector switch to zero.
4.2 Obtaining an AC voltage
•
Connect the load to the AC output sockets (3).
•
Set the voltage selector switch (3) to give the
required voltage; if necessary connect a voltme-
ter in parallel with the load.
4.3 Obtaining a DC voltage
•
Connect the load to the DC output sockets (1 or 2).
•
Set the voltage selector switch (1 or ) to give the
required voltage; if necessary connect a voltme-
ter in parallel with the load.
5. Sample experiments
5.1 External photoelectric effect (Hallwachs ef-
fect)
Equipment
1 DC Power Supply 450 V (230 V, 50/60 Hz)
U8521400-230
1 Control Unit for Spectrum Lamps (230 V, 50/60 Hz)
U21905-230
1 Electrometer (230 V, 50/60 Hz)
U8531408-230
or
1 DC Power Supply 450 V (115 V, 50/60 Hz)
U8521400-115
1 Control Unit for Spectrum Lamps 115 V, 50/60 Hz)
U21905-115
1 Electrometer (115 V, 50/60 Hz)
U8531408-115
1 High-Pressure Mercury Vapour Lamp
U8473155
1 Grid electrode, 1 Zinc electrode,
1 Metal rod with 4-mm drilled hole
1 SK plug-in resistor 10 G
Ω
from Electrometer Accessories
U8531420
1 Multimeter AM50
U17450
1 Tripod Stand, 150 mm
U13270
1 Stainless Steel Rod, 250 mm
U15001
1 Universal Clamp
U13255
Connection leads
5.2 Recording equipotential lines of electric
fields
Equipment
1 DC Power Supply 450 V (230 V, 50/60 Hz)
U8521400-230
or
1 DC Power Supply 450 V (115 V, 50/60 Hz)
U8521400-115
1 Electrolyte Trough
U51001
1 Multimeter AM50
U17450
Connection leads
400 cm
3
distilled water