12.3 Assembling the equipment:

In case castor oil had been removed from the oil
chamber:

•

Pour a fresh quantity of castor oil in up to about
5 mm below the upper edge of the cylinder (at
the beginning of the depression).

•

Insert both of the silicone seals.

•

Turn the conical seal inside out and dampen the
stud with some castor oil then screw it into the
guide tube.

•

Unfold the conical seal back to its original shape,
position the spring on the piston and insert the
guide tube into the piston.

•

Mount the measuring cell and position it flush
along the edges of the cylinder.

•

Place the heat casing at the centre of the lower
silicone seal.

•

Fit the circular grommet and, with the help of a
ruler placed on the heat casing, position it paral-
lel to the cylinder (see Fig. 9, the semicircular
holes should then be below the valve openings).

Fig. 9: Positioning the circular grommet

•

Place the valve plate at the centre and position it
parallel to the end plate.

•

Fit the M8

×40 screws with new copper gasket

washers and loosely screw them in.

•

Tighten the screws. Take care to ensure that there
is uniform pressure on the circular grommet (if
the pressure is too high, the grommet makes a
greyish mark on the transparent acrylic, whereas
if the pressure is lower the surface looks milky).

12.4 Recommissioning:

•

Degas the hydraulic fluid and pour the oil into
the equipment (see chapter 10).

•

Position the safety valve (see chapter 10).

•

Conduct a fresh volume calibration (see chap-
ter 6).