3: Operation

12

041-00000-010-02-201106

5. Flush the LWCC with de-ionized water or experimental buffer solution using a pump or a syringe

at a rate of 1mL/minute and observe the light intensity or absorbance baseline on the detector.
Continue flushing until the signal is stable. (See

Troubleshooting

in Chapter 4:

Maintenance

if

the signal does not stabilize appropriately.)

Measuring Discrete Samples

Discrete samples can be measured with the LWCC by introducing the sample with a syringe. Sample
volumes approximately 1.5 – 3 times the cell volume are necessary to fill the LWCC. When injecting a
sample or reference solution, apply consistent hand pressure to the syringe. Small variations in baseline
levels may result from using this method.

►

Procedure

1. Using fiber optic cables, connect the LWCC to a light source and a detector.

2. Clean the LWCC using the standard cleaning procedure described in Chapter 4:

Maintenance

.

3. Flush the LWCC with de-ionized water or experimental buffer solution using a syringe and

observe the light intensity or absorbance baseline on the detector until the signal is stable.

4. Introduce the sample slowly into the LWCC with a syringe using steady pressure to avoid

generating air bubbles.