Ocean Optics HR2000CG-UV-NIR User Manual
Page 39

Appendix A
Saving the New Calibration Coefficients: Serial Mode
If you are connecting the HR2000 Spectrometer to the serial port of the PC, you need to save the new wavelength
calibration coefficients to the .SPEC file that OOIBase32 accesses when opened.
Note: You cannot save the calibration coefficients to the EEPROM memory chip on the HR2000 when using the
serial mode.
To Save Wavelength Calibration Coefficients Using the Serial Mode
1. Open the OOIBase32 application.
2. Select
Spectrometer | Configure from the OOIBase32 menu bar. The Configure Spectrometer screen
appears.
3. Select
the
Wavelength Calibration tab to update the wavelength coefficients within OOIBase32.
4. Enter in the new values acquired from Steps 5 and 6 of the Calibrating the Wavelength of the
Spectrometer section in this Appendix.
HR2000 High-Resolution Fiber Optic Spectrometer
35
5. Click
the
OK button to save the information in OOIBase32.