Performing a user calibration with spectrasuite – Ocean Optics pH Sensor Patches, Probes and Cuvettes User Manual
Page 20

2: Desktop Software Installation and Operation
14
PHSENSOR-TRANS-REFLECT-201111
12. You are now ready to take pH measurements. See
Taking pH Measurements with SpectraSuite
.
Performing a User Calibration with SpectraSuite
► Procedure
1. Open SpectraSuite and select File | New | New Sol Gel pH Measurement.
2. Click the Calibration Wizard button to begin the calibration.
3. Select the spectrometer to use and click Next.
4. Select Perform Independent Calibration and then click Next. The Enter Experimental
Parameters screen appears.
5. Enter your Experimental parameters: Acquisition Wavelength, Baseline Wavelength, and
approximate Ambient Temperature. Click Set, then click Next.
Note
For pH sensors that perform in the biological range (pH 5 – 9), the Acquisition Wavelength is
620nm and the Baseline Wavelength is referred to in the above specification tables for the
various form factors.