Laser troubleshooting, Problem, Probable cause – Ocean Optics LIBS2000+ User Manual
Page 43: Solution, Spectrometer troubleshooting, Continuous beeping from system

4: Troubleshooting
Laser Troubleshooting
Problem
The laser will not start.
Probable Cause
On water-cooled lasers, bubbles may have developed in the cooling system. This will prevent water from
traveling to the laser head and result in a nonfunctional laser.
Solution
►
Procedure
1. Tilt the power supply of the laser 30 degrees right or left.
2. Allow the power supply to remain in that position until the bubbles clear.
3. Try to start the laser again.
Spectrometer Troubleshooting
Continuous Beeping from System
Problem
The system is continuously beeping.
Probable Cause
The PC cannot communicate with the USB devices.
Solution
►
Procedure
1. Remove the USB devices from the PC,
2. Remove all installed Ocean Optics USB drivers, and
3. Restart the system.
4. Ensure that OOILIBS is installed, then reconnect the LIBS2000+ Spectrometer to the PC.