Bulb replacement, Overview, Hl-2000-fhsa-hp – Ocean Optics OTS User Manual
Page 19: Replacing the bulb, Appendix a: bulb replacement, Appendix a

Appendix A
Bulb Replacement
Overview
The OTS system contains the HL2000-HP-FSHA light source, which consists of a halogen bulb (Part No.
HL-2000-HP-B), attenuator and TTL shutter. Specifications for this light source are as follows:
HL-2000-FHSA-HP
Wavelength range
360 nm – 1700 nm
Stability 0.5
%
Drift
<0.1% per hour
Time to stabilize
Approximately 5 Minutes
Output to bulb
2V DC / 1,67A
Bulb life time
2000 h
Characteristic Focused
Shutter
TTL max. 5Hz
DB-15 Connector
PIN 13: TTL PIN 10: Ground
Bulb color temperature
3.000K
Room temperature
5°C – 35°C
Humidity
5 - 95% at 40°C
Output 20W
WARNING
Before replacing the bulb in the HL2000-HP-FHSA, disconnect the lamp from your
power source and allow the unit to cool for at least twenty minutes, if necessary.
Replacing the Bulb
►
Procedure
Refer to the following figure and perform the steps below to replace the bulb in the HL-2000-
FHSA Light Source:
OTS-00000-000-02-1008
13