Bulb replacement, Overview, Hl-2000-fhsa-hp – Ocean Optics OTS User Manual

Page 19: Replacing the bulb, Appendix a: bulb replacement, Appendix a

Advertising
background image

Appendix A

Bulb Replacement

Overview

The OTS system contains the HL2000-HP-FSHA light source, which consists of a halogen bulb (Part No.
HL-2000-HP-B), attenuator and TTL shutter. Specifications for this light source are as follows:

HL-2000-FHSA-HP

Wavelength range

360 nm – 1700 nm

Stability 0.5

%

Drift

<0.1% per hour

Time to stabilize

Approximately 5 Minutes

Output to bulb

2V DC / 1,67A

Bulb life time

2000 h

Characteristic Focused
Shutter

TTL max. 5Hz

DB-15 Connector

PIN 13: TTL PIN 10: Ground

Bulb color temperature

3.000K

Room temperature

5°C – 35°C

Humidity

5 - 95% at 40°C

Output 20W

WARNING

Before replacing the bulb in the HL2000-HP-FHSA, disconnect the lamp from your
power source and allow the unit to cool for at least twenty minutes, if necessary.

Replacing the Bulb

►

Procedure

Refer to the following figure and perform the steps below to replace the bulb in the HL-2000-
FHSA Light Source:

OTS-00000-000-02-1008