Port, consult the, Saving the new calibration, Coefficients: serial mode – Ocean Optics HR2000+ Install User Manual
Page 23: Section later

A: Calibrating the Wavelength of the HR2000+
294-00000-000-02-201012
17
Saving the New Calibration Coefficients: Serial
Mode
If you are connecting the HR2000+ Spectrometer to the serial port of the PC, you need to save the new
wavelength calibration coefficients to the .SPEC file that OOIBase32 accesses when opened.
Note
You cannot save the calibration coefficients to the EEPROM memory chip on the
HR2000+ when using the serial mode.
►
Procedure
To save Wavelength Calibration Coefficients using the Serial mode, perform the following steps:
1. Open the OOIBase32 application.
2. Select Spectrometer | Configure from the OOIBase32 menu bar. The Configure Spectrometer
screen appears.
3. Select the Wavelength Calibration tab to update the wavelength coefficients within OOIBase32.
4. Enter in the new values acquired from Steps 5 and 6 of the
Calibrating the Wavelength of the
section in this appendix.
5. Click the OK button to save the information in OOIBase32.