3: R-LS-450 Rack-mount Setup and Operation

009-00000-450-02-201105

3

The following table contains information on the function of each set of pins in JP2.

 Jumpering the CW pins makes the R-LS-450 operate continuously. This means that there is no

pulsing of the light source. Other jumper blocks must be configured correctly for this setting to
work properly.

 Jumpering sets 2

16

, 2

15

, 2

14

, 2

13

, 2

12

, 2

11

, and 2

10

controls the pulse rate per second of the R-LS-

450, depending on the A/D converter you are using to interface to the spectrometer (see the table
below for pulse rates).

 Jumpering the CS pins controls the pulse rate via the spectrometer’s Continuous Strobe Setting in

OOIBase32 software (see the Using JP3 on the S2000 section that follows for more information).

JP2 Pins

Function

ADC1000 Frequency (Hz)

ADC2000-PCI Frequency (Hz)

CW

Continuous mode

0

0

2

16

Divide by 2

16

15.2

30.4

2

15

Divide by 2

15

30.4

60.8

2

14

Divide by 2

14

60.8

122.0

2

13

Divide by 2

13

122.0

244.0

2

12

Divide by 2

12

244.0

488.0

2

11

Divide by 2

11

488.0

976.0

2

10

Divide by 2

10

976.0

1952.0

CS

Continuous Strobe

Software-controlled

Software-controlled

Jumper Block 3 (JP3)

JP3 consists of two sets of pins. The jumper position determines the source of power control (manual or
remote) for the R-LS-450. Jumping the Remote pins allows you to control the R-LS-450 through the
software (if other jumper blocks are configured correctly).

R-LS-450 Operation Matrix

Use the following operation matrix to help you determine the correct jumper settings:

S1 Switch

JP1

JP3

LED Status

Off

No jumper

No jumper

Off

CW

No jumper

No jumper

Continuously On