Ocean Optics MultiFrequency Phase Fluorometer User Manual
Page 24

3: Configuration
2. Set the MFPF parameters (located in the Instrument Controls box on the lower left of the main
screen) according to the following recommendations:
Parameter Setting
Dual Channel
Unchecked for set up, checked if 2 sensors are used
Sampling Period
2 seconds for set up
Frequency kHz
5 kHz – FOSPOR
40 kHz – FOXY and HIOXY
APD voltage
110 to 155 V photon-to-electron ratio
LED Channel
Depends on your experiment
Detector Channel
Depends on your experiment
3. Select Advanced | Show More Control Panel from the menu bar. The following dialog box
appears:
4. Set the parameters according to the following recommendations:
Parameter Setting
Analog AGC
Checked
Digital AGC
Checked
LED Mode
Intermittent
LED Percentage
25 -- 100%