5: Operating the JAZ-INDY Module
013-RD000-000-02-201103
41
Menu Item
Description
Module Info
Allows you to display the Indy module serial number where:
v = the CCA revision
xxxx = the 4-digit number of the device
Calibrate
Select to measure the following:
voltage out: in counts or volts
voltage in: in counts or volts
4 – 20 mA Transmit
4 – 20 mA Receive
See
Calibrating the JAZ-INDY Module
for the calibration procedures.
JAZ-INDY Menu Defaults
When you first power-on the Jaz stack, the JAZ-INDY module defaults to the following settings:
Setting
Default Value
GPIO Menu
Set Direction
1-8: Input
Set Output Values does not appear until one of the 1
bits is changed to Output
Analog V Out
V Output
- 5.00
4-20mA Excitation
4-20mA Excitation
10 Volts (+)
4-20mA Tx Control
Set Value
.0
Calibrating the JAZ-INDY Module
Your JAZ-INDY module is calibrated at the factory, but you can use the procedures in this section to
recalibrate the module so that it matches your equipment.
You can calibrate the following on the JAZ-INDY module:
Four voltage outputs