Ocean Optics XYZ Mapping Tables User Manual
Page 19

3: Operating Instructions
000-10000-130-02-1209
11
Commands for Evaluating Limit Switches and Homing Points (Cont’d)
Command
Description
Example
HB*
Hard blocking
Activates hard blocking for given limit switch
1**: activate
0**: deactivate
HB3
HD*
Hard direction
Sets the direction to be blocked
1**: right rotation blocked
0**: left rotation blocked
HD2
GOHIX
Go hall index
Motor runs to hall zero and sets the position to 0
GOHIX
HS
Hard status
A number 1-3 indicating which limit switches have activated
HS
-> 3
GAHS
Get actual homing status
5 numbers with values between 0 and 3 are sent to the host. Indication the
status of the limit switches
HA value
HL
HN
HB
HD
GAHS
-> 33300
GHSC
Get homing sequence configuration
3 numbers between 0 and 3. Indicating the settings of the homing
sequence
HA
HL
HN
GHSC
-> 220
Commands for Parameters
Command
Description
Example
SP* ***
Load maximum speed
Loads new maximum velocity (rpm) Range 0 to 30000
SP1000
GSP
Get maximum speed
Calls up actual maximum velocity
GSP
-> 1000