Advanced instruments inc – Analytical Industries GPR-1900 Series Trace PPB Oxygen Analyzer User Manual

Page 24

Advertising
background image

Advanced Instruments Inc.



24

0.000 PPM

ZERO
CALIBRTION
ENTER TO CALIBRATE
MENU TO ABORT


8. Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode.

9. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or

reading has stabilized within 50% of the full scale low range.

10. Both the Zero Calibrate and Span Calibrate functions result in the following displays:



PASSED

CALIBRATION

OR


FAILED

CALIBRATION





Default Zero:
The software will eliminate any previous zero calibration adjustment and display the actual the signal output of the sensor at
any specific oxygen concentration. For example, assuming a zero gas is introduced, the display will reflect an oxygen reading
representing basically the zero calibration adjustment as described above. This feature allows the user to test the sensor’s signal
output without removing it from the sensor housing.

Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:

MAIN MENU

AUTO SAMPLE

MANUAL SAMPLE

CALIBRATION

CONFIG ALARMS
BYPASS ALARMS



>>>

CALIBRATION

SPAN CALIBRATE

ZERO CALIBRATE
DEFAULT SPAN

DEFAULT ZERO

OUTPUT SPAN
OUTPUT ZERO

5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears and after 3 seconds the system returns to the SAMPLING mode:



FACTORY

DEFAULTS

SET




3.3

PPM

AUTO SAMPLING
100 PPM RANGE

24.5 C
LO1 2 PPM 10 PPM HI2