Advanced instruments inc – Analytical Industries GPR-1600 MS Series PPB Oxygen Analyzer User Manual

Page 31

Advertising
background image

Advanced Instruments Inc.



31

CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE

DEFAULT SPAN

DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO


11. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
12. Press the ENTER key to select the highlighted menu option.
13. The following display (below left) appears and after 3 seconds the system returns to the SAMPLING mode:



FACTORY

DEFAULTS

SET




3.3

PPM

AUTO SAMPLING

10 PPM RANGE

24.5 C
LO1 2 PPM 10 PPM HI2


14. Repeat steps 1 through 4 above.

CALIBRATION
SPAN CALIBRATE

ZERO CALIBRATE

DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN

OUTPUT ZERO

15. Advance the reverse shade cursor using the ARROW keys to highlight ZERO CALIBRATE.
16. Press the ENTER key to select the highlighted menu option.
17. The following displays appear:

0.000 PPM


ZERO
CALIBRATION
ENTER TO CALIBRATE
MENU TO ABORT



18. Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode.
19. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or

reading has stabilized within 50% of the full scale low range.

20. Both the Zero Calibrate and Span Calibrate functions result in the following displays:



PASSED

CALIBRATION

OR


FAILED

CALIBRATION