11. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.

12. Press the ENTER key to select the highlighted menu option.
13. The following display appears:

100.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN



14. Enter the calculated adjustment value. Note: Once the initial adjustment is made and checked at the PLC it may be

necessary to fine tune the initial adjustment by repeating. Any additional percent error must be added or subtracted from
the initial adjustment value

000.0

OUTPUT ZERO OFFSET

PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN





15. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to

reach to the desired digit of the adjustment OUTPUT ZERO OFFSET value.

16. Press the ARROW keys to enter each the numerical value of each digit of the adjustment OUTPUT ZERO OFFSET value.
17. Repeat until the complete OUTPUT ZERO OFFSET value has been entered.
18. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
19. The system returns to the SAMPLING mode.

Output Span
20. Access the MAIN MENU by pressing the MENU key.
21. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
22. Press the ENTER key to select the highlighted menu option.
23. The following displays appear:

MAIN MENU

AUTO SAMPLE
MANUAL SAMPLE

CALIBRATION



>>>

CALIBRATION

SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO

OUTPUT SPAN

OUTPUT ZERO



24. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT SPAN.
25. Press the ENTER key to select the highlighted menu option.
26. The following display appears: