7 Troubleshooting


If the recommended corrective action does not resolve the problem return the
device to the factory for service.

Symptom

Corrective Action

Device appears to be physi-
cally damaged

Turn device ON – if it successful passes
calibration – proceed

No display when analyzer is
turned ON

Replace batteries
Check battery polarity
Check and/or clean battery contacts

‘LO’ displayed when analyzer is
turned ON or in use.

Replace battery and calibrate device

‘ERR’ displayed when analyzer
fails calibration

Replace oxygen sensor

No response to keypad

Replace battery

Cannot turn device OFF

Calibration routine in process – wait until
completed

Reading displayed by LCD
does not change when oxygen
level changes

Replace sensor

Reading does not stabilize or
fluctuates erratically

Relocate analyzer away source of RF or
electromagnetic radiation emissions.
Wait 5 minutes and repeat calibration
Replace sensor, repeat calibration

After calibration in 100% dry
oxygen, analyzer reading drifts
more than 2% over 8 hours

Check primary oxygen delivery device
Replace sensor that is nearing the end of
its useful life

8 Specifications

Accuracy:

< 2% of FS range under constant conditions

Analysis: 0-100%

oxygen

Alarms: Analyzer

none

Calibration:

Certified dry 100% oxygen or air after 8 hrs of use

Compensation: Temperature

Connections:

1x16mm thread or o-ring diverter

Controls:

Soft touch keypad for ON/OFF and CAL

Dimensions:

2.72” x 4.1” x 1.35”; weight 7 oz. (196 grams)

Display:

3 digit LCD 1.1” x .625”; resolution 0.1% O

2

Flow Sensitivity:

None between 0.2 to 10 liters per minute

Humidity: Non-condensing

0-95%

RH

Linearity: + 1% under constant conditions

Pressure:

Inlet – ambient or regulated; vent - atmospheric

Power:

(2) 1.5V AA alkaline batteries; 13,000 hrs of use

Response Time:

90% of final FS reading in 10 seconds

Sensitivity:

< 0.5% of FS range

Sensor:

AII-11-75-PO2D; optional remote sensor AII-11-75-PO2RD

Sensor Life:

32 months in air at 25ºC and 1 atmosphere

Storage Temp.:

-20º to 60ºC (-4ºF to 140ºF) on intermittent basis

Temp. Range:

5º to 45ºC (41ºF to 113ºF)

Warm-up Time:

None

Warranty:

12 months analyzer; 12 months sensor (any application)