Specifications – YSI 55 User Manual

Page 6

Advertising
background image

2

2. Specifications

Probe Operating Environment

Medium: fresh, sea, or polluted water

Temperature: -5 to +45

o

C

Depth: 0 to 12, 0 to 25 or 0 to 50 feet (depending on cable length)

Meter Ambient Operating/Storage Temperature: -10 to +50

o

C

Material: ABS, Stainless Steel, Acrylic, and other materials.

Dimensions:

Height:

9.5

inches

(24.13

cm)

Thickness:

2.2

inches

(5.6

cm)

Width:

3.5 inches max.

( 8.89 cm)

Weight:

1.7 pounds

( 0.77 kg)

Power: 9 VDC - 6 AA-size Alkaline Batteries (included)

Approximately 100 hours operation from each new set of batteries

Water Tightness: Meets or exceeds IP65 standards

Extensive testing of the YSI Model 55 suggests the following typical performance:

Temperature

Sensor Type................Thermistor
Range..........................-5 to +45

o

C

Accuracy ....................± 0.2

o

C

Resolution ..................0.1

o

C

Dissolved Oxygen % Saturation

Sensor Type................Membrane covered polarographic
Range..........................0 to 200 % air saturation
Accuracy ....................± 2 % air saturation
Resolution ..................0.1 % air saturation

Dissolved Oxygen mg/L

Sensor Type................Calculated from % air saturation, temperature and salinity.
Range..........................0 to 20 mg/L
Accuracy ....................± 0.3 mg/L
Resolution ..................0.01 mg/L

Advertising