The lcd display, Operational procedures, Measurement modes – YSI DO200 User Manual

Page 6: Calibration set-up, Requirements

Advertising
background image

THE LCD DISPLAY

A

BA

T

SAL ppt

mBar

% ppm

1

2

3

4

5

7

C

CAL

6

Figure2. LCD Display

1. BAT: Low battery indicator.

2. CAL: Calibration mode indicator.

3. SAL

ppt: Displays during calibration

when user is prompted for the
approximate salinity of the sample in
parts per thousand (ppt).

4. mBar: Displays during calibration to

prompt user for barometric pressure.

5.

Main display for dissolved oxygen

values.

6. %/ppm: Unit indicators.

7. ºC: Temperature display.

OPERATIONAL PROCEDURES

Press

to turn the unit on or off. The instrument will perform a self-diagnostic test, during which

an “ovEr” message may appear on the display. This is normal. After the warm up is complete
the message will disappear. After the self-diagnostic test completes, the temperature displays in
the lower right of the display, and the unit is ready for operation. Immerse the probe halfway into
the sample solution. If possible, do not allow probe to touch any solid object in the solution.
Allow no air bubbles around the probe. When the unit is not in use, turn it off to save battery life.

NOTE: During an oxygen measurement, the probe must be moved approximately 1/2 ft per
second to overcome the inherent consumption of oxygen by the sensor. When using the 200-
BOD lab probe, however, simply use the probes self-stirring feature.

MEASUREMENT MODES

This unit provides three distinct measurements:

1. Temperature - Current solution temperature continually displays.

2. Dissolved

Oxygen

% - Measurement of oxygen in percent saturation.

3. Dissolved

Oxygen

ppm - Measurement of oxygen in ppm (mg/L).

Carefully observe the units displayed at the far side of the LCD to determine the desired mode.

CALIBRATION SET-UP

Requirements

1. The approximate pressure (in millibars [mBar]) of the region to be measured for dissolved

oxygen.

2. The approximate salinity of the water to be analyzed. Fresh water has an approximate

salinity of zero. Seawater has an approximate salinity of 35 parts per thousand (ppt).

3. For highest accuracy, complete all calibrations at a temperature as close as possible to the

sample temperature.

5

Advertising