TROUBLESHOOTING

Main Display reads:

Possible Solutions:

“ovEr” or “undr”

• Check membrane and electrolyte solution.
• Clean anode and cathode.
• Return product for service.

Secondary Display reads:

Possible Solutions:

“undr”

• Heat the sample to above –6.0 °C
• Return product for service.

“ovEr”

• Cool sample to below 46.0 °C
• Return product for service.

SPECIFICATIONS

Display Range

Accuracy

Resolution

Dissolved O

2

(ppm or mg/L)

0 to 20.00 ppm (mg/L)

±2 % of the reading or ±2% air

saturation, whichever is greater

0.01 mg/L

Dissolved O

2

% air-sat

0 to 200.0 %

±2% of the reading or ±0.2 ppm,

whichever is greater.

0.1 %

Temperature °C

-6.0 to 46.0 °C

21 to 115 °F

±0.3 °C ±1 digit

0.1 °C

Pressure Compensation

600 to 1100 mBar (450 to 825 mmHg)

Salinity Compensation

From 0.0 to 40.0 ppt

ATC Probe

Thermistor, 10K

Ω, at 25°C

Calibration Backup

Yes

Audio Feedback

Yes, on all keys

Power Source

One 9V battery

Operating Temperature

0 to 50

°C (32 to

122 °F)

Instrument Casing

Splash-resistant IP 65

Weight (with battery)

350 grams (.75 lbs)

Dimensions (L x W x D)

186 mm x 70 mm x 37 mm (7.3 in x 2.8 in x 1.5 in)

CONVERSIONS

To Convert:

Multiply by:

Inches of Hg to mBar

33.864

Inches of Hg to mmHg

25.4

mmHg to mBar

1.333

RECOMMENDED SPARE PARTS LIST

PART #

DESCRIPTION

200-4

4 meter (approx. 13 feet) probe and cable assembly

200-10

10 meter (approx. 33 feet) probe and cable assembly

200-BOD

Self-stirring BOD lab probe and cable assembly with power supply

485

DO carrying case, soft sided

280

DO carrying case, hard sided

5908

Membrane kit, 1.25 mil PE (605306), six cap membranes and KCl solution

480

Instrument carrying case, soft


Item #605368 • Drawing #A605368

Revision E • March 2008