Probe maintenance, Troubleshooting, Main display – YSI EC300 User Manual

Page 7: Problem, Possible solution, Over, Over/undr during calibration, Secondary display, Over/undr, Overundr

Advertising
background image

2. Press

MODE to enter the desired measurement mode. The message ‘rAng’ (range) may

appear briefly on the display indicate auto-ranging; this is normal. Allow temperature to
stabilize before taking measurements.

PROBE MAINTENANCE

The most important requirement for accurate and reproducible conductivity measurements is a
clean cell. A dirty cell changes the conductivity of a solution through contamination. Clean the
cell thoroughly before storing it. To clean the conductivity cell:


1.

Dip the cell in cleaning solution and agitate for two to
three minutes. Any foaming acid tile cleaner, such as
Dow Chemical Bathroom Cleaner, should clean
adequately. For a stronger cleaner, use a solution of
1:1 isopropyl alcohol and 1 N HCl. Remove the cell
from the cleaning solution.


2.

Use the nylon brush (supplied) to dislodge any
contaminants from inside the electrode chamber.


3.

Repeat steps one and two until the cell is completely clean. Rinse the cell thoroughly in
deionized, or clean tap water.

TROUBLESHOOTING

MAIN DISPLAY

PROBLEM

POSSIBLE SOLUTION

OvEr

• Conductivity is >200.0 mS

• Salinity is > 70.00 ppt

OvEr/Undr during calibration

Cell Constant Calibration is

out of range

• Completely submerge the

probe.

• Allow sufficient time for the

electrode and Temp probe
stabilization.

• Recalibrate with correct value

for the conductivity standard.

• Replace conductivity standard.

• Clean cell.

• Return for service.

MAIN

DISPLAY

SECONDARY

DISPLAY

OvEr/Undr


OvEr

Undr

Temperature >90.0

°C

Temperature < -10.0

°C

• Decrease/Increase the sample

temperature.

• Return for service.

6