Specifications, Recommended spare parts list – YSI EC300 User Manual

Page 8

Advertising
background image

7

SPECIFICATIONS

Display Range

Accuracy

Resolution

Conductivity,
Auto-ranging

0.0 to 499.9 uS/cm
500 to 4999 uS/cm
5.00 to 49.99 mS/cm
50.0 to 200.0 mS/cm

±1% of reading plus 2 uS/cm
±1% of reading plus 5 uS/cm

±1% of reading plus 0.05 uS/cm

±2.5% of reading plus 0.5 mS/cm

0.01 uS/cm

1 mS/cm

0.01 mS/cm

0.1 mS/cm

Salinity

0.0 to 70.0 ppt

0.2% Full Scale

0.1 ppt

Temperature °C

-10.0 to 90 °C

±0.2 °C or ±0.4% Full Scale,

whichever is greater

0.1 °C

Reference Temperature

15.0 to 25.0

°C

Temperature Coefficient

0.0% to 4.0%

Cell Constant

5.00

±

0.50

TDS Constant Range

0.30 to 1.00

Power

One 9V battery

Calibration Back-up

Yes

Audio Feedback

Yes, on all touch keys

Water Resistance

Splash-resistant, IP 65

Operating Temp. Range

0 to 50

°C

Operating Relative Humidity Range

up to 95%

Temperature Probe

Thermistor, 10k

Ω / 25

°C

Dimensions (L x W x D)

186 mm x 70 mm x 37 mm (7.3 in x 2.8 in x 1.5 in)

Weight (batteries included)

430 grams (1 lb)

RECOMMENDED SPARE PARTS LIST

PART #

DESCRIPTION

300-4

4-meter probe and cable assembly.

300-10

10-meter probe and cable assembly.

380

Carrying case, hard sided.

480

Instrument carrying case with shoulder strap, soft sided.














Item #605369 • Drawing #A605369

Revision D • March 2008

For the most recent version of this manual, visit www.ysiecosense.com

Advertising