6 function check, Function check -5 – YSI IQ S ENSOR N ET MIQ/CHV PLUS User Manual

Page 15

Advertising
background image

MIQ/CHV PLUS

Installation

3 - 5

ba76036e01

01/2012

3.6

Function check

General instructions

This check is also suitable for the troubleshooting of malfunctions.

If the valve is not controlled via the IQ S

ENSOR

N

ET

system, the valve

control line must be connected.

To check that the valve triggers correctly, proceed as follows:

Warning

Danger of injury from lines that are under pressure. The
compressed air glands in the housing may only be opened by a
service technician authorized by YSI.

1

MIQ/CHV PLUS + IQ S

ENSOR

N

ET

:

Put the IQ S

ENSOR

N

ET

system into operation.

MIQ/CHV PLUS + measuring transmitter:
Supply the power supply module with voltage and put the
measuring transmitter into operation.

2

Adjust the cleaning intervals on the measuring system so that
the valve opens.
or
Open the valve in the Valve function menu with the Manual
control
function (see section 4.4.3).

3

Check whether there is a compressed air stream on the MIQ/
CHV PLUS.