2 sensor-controlled, Ee section 4.4.2 – YSI IQ S ENSOR N ET MIQ/CHV PLUS User Manual

Page 25

Advertising
background image

MIQ/CHV PLUS

Settings

4 - 9

ba76036e01

01/2012

Canceling the cleaning

A running cleaning cycle is canceled:

Automatically

– If the sensor switches to the inactive condition during the cleaning

cycle

Manually

– By pressing

c

– By switching on the maintenance condition

Each time the cleaning cycle is canceled, the valve closes immediately.

If the cleaning cycle is canceled automatically, the outputs linked to the
sensor are released immediately.
If the cleaning cycle is canceled manually, the sensor is in the
maintenance condition. The linked outputs are only released after the
maintenance condition was terminated manually.

The next cleaning cycle will be performed at the time set up.

Note
The valve closes in the case of power failure. The cleaning cycle is
canceled. As soon as power is available again the next cleaning cycle
takes place at the specified time.

4.4.2

Sensor-controlled

With the Sensor-controlled function, the valve is controlled by a linked
sensor. A sensor (or connection module) is required that sends signals
to prompt a cleaning cycle, e.g. the MIQ/VIS connection module for UV/
VIS sensors.

Note
Set the cleaning process in the setting table of the respective sensor.

Advertising